Support

Lorem ipsum dolor sit amet:

24h / 365days

We offer support for our customers

Mon - Fri 8:00am - 5:00pm (GMT +1)

Get in touch

Cybersteel Inc.
376-293 City Road, Suite 600
San Francisco, CA 94102

Have any questions?
+44 1234 567 890

Drop us a line
info@yourdomain.com

About us

Lorem ipsum dolor sit amet, consectetuer adipiscing elit.

Aenean commodo ligula eget dolor. Aenean massa. Cum sociis natoque penatibus et magnis dis parturient montes, nascetur ridiculus mus. Donec quam felis, ultricies nec.


Bottle-top burette Titrette®

Bottle-top burette Titrette®

for quick and reliable titration with highest precision and with no power hookup needed, observing the following physical limits:

  • +15 °C to +40 °C (59 °F to 104 °F) of instrument and reagent
  • vapor pressure up to 500 mbar
  • viscosity up to 500 mm2/s
  • altitude: maximum 3000 m above sea level
  • relative humidity: 20 % to 90 %
  • within the error limits of class A
  • digital display
  • Easy Handling (no switching between „fill“ and „titrate“)
  • simple maintenance (cleaning and part replacement right in the lab)
  • optionally available with PC interface
  • four additional electronic functions:
    • adjustment with Easy Calibration
    • calibration schedule
    • save power with Auto-Power-Off
    • changing decimal place settings
  • with bottle thread GL 45
  • dismountable
  • titrating tube horizontally and vertically adjustable for flexible positioning
  • DE-M marking
  • items supplied:
    • bottle-top burette Titrette®
    • telescoping filling tube 170-330 mm
    • recirculation tube
    • 2 batteries (AAA/UM4/LR03)
    • 3 polypropylene bottle adapters (GL 45/32, GL 45/S 40, GL 32/NS 29/32)
    • 2 colored light shield inspection windows
    • operation manual
    • performance certificate
Cat. no. Volume Resolution UE
5334021 10 ml 0.001 ml 1
5334023 25 ml 0.001 ml, above 20 ml titration volume 0.01 ml 1
5334024 50 ml 0.002 ml, above 20 ml titration volume 0.01 ml 1

Error limits

Volume Partial volume A* ≤ ± % µl CV* ≤ % µl
10 ml 10 ml
5 ml
1 ml
0.10
0.20
1.00
10
10
10
0.05
0.10
0.50
5
5
5
25 ml 25 ml
12.5 ml
2.5 ml
0.07
0.14
0.70
18
18
18
0.025
0.05
0.25
6
6
6
50 ml 50 ml
25 ml
5 ml
0.06
0.12
0.06
30
30
30
0.02
0.04
0.20
10
10
10

*(A = accuracy, VK = coefficient of variation)

Error limits related to the nominal capacity (= maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at ambient temperature (20°C/68°F) and with smooth operation.

The instrument can be used for the following titration media (maximum contentration 1 ml/l):

Acetic acid Iron (II) sulfate solution Silver nitrate solution*
Alcoholic potassium
hydroxide solution
Nitric acid Sodium arsenite solution
Ammonium iron (II)
sulfate solution
Oxalic acid solution Sodium carbonate solution
Ammonium thiocyanate solution Perchloric acid Sodium chloride solution
Barium chloride solution Perchloric acid in glacial
acetic acid
Sodium hydroxide solution
Bromide bromate solution Potassium bromate solution Sodium nitrite solution
Cerium(IV)sulfate solution Potassium bromide bromate
solution
Sodium thiosulfate solution
EDTA solution Potassium dichromate solution Sulfuric acid
Hydrochloric acid Potassium hydroxide solution Tetra-n-butylammonium
hydroxide solution
Hydrochloric acid in acetone Potassium iodate solution Triethanolamine in acetone*
Iodide iodate solution* Potassium permanganate solution* Zinc sulfate solution
Iodine solution* Potassium thiocyanate solution  

*use light shield inspection window

When the instrument is properly handled, dispensed liquid will only come into contact with the following chemically resistant materials:

borosilicate glass, Al2O3, ETFE, PFA, FEP, PTFE, platinum-iridium, PP (screw cap)

Copyright 2024 - Paul Marienfeld GmbH & Co. KG